ETi-9110
-
Features
Analysis Range: 100ppm ~ 1000 ppm
*The patent design of both of quantitative sampling and dilution device require no maintenance and regular flow calibration.
*The use of large caliber sampling cup and sampling tube causes no pipe blockage.
The sampling tube and drain tube with large aperture pneumatic valve, which prevents from the fouling or blockage and it may happened in traditional solenoid valve.
*Use the Titration Curve Differentiation to find the titration end point automatically in order to avoid the misjudgment of titration end point because of electrode offset potential.
*The detection electrode is copper electrode and its probe is a solid chip (with one year above of service life)
The probe of calcium electrode is PVC membrane whose service life is about 3 ~ 6 months (it depends on the water quality).
*Chinese graphic operation interface with easy operation, which can analyze data and search for the historical data and alarm message.
*Auto regular calibration (the cycle time could be set by program).
It can manually analyze the foreign sample for the person in laboratory to do the comparison conveniently.
内容说明
Model. |
ETi-9110 |
Display Screen |
5.7 inches of color touch screen |
Analysis Range |
100 ~ 1000 ppm |
pH Value of Sample Water |
Range: PH3 ~ PH10 |
Sample Temperature |
5 ~ 45℃ |
Analysis Cycle Time |
6~15mins.(depends on its concentration) |
Accuracy |
1000 ± 70 ppm. |
Reproducibility |
± 3 % full scale |
Resolution |
0.1 ppm |
AutoCorrect Message and Calibration Record |
Auto Updating is not required. |
Standby Time |
0 ~ 7200 sec |
Historical Data and Record |
V |
Alarm Message and Record |
V |
Liquid Level Detection |
V |
Auto dosing |
V |