7714

内容说明

 Conductivity Measure Unit Specification

 

Model

7714

Measurement range

0~4μS/cm (0~0.4mS/m)

0~20μS/cm (0~2mS/m)

0~400μS/cm (0~40mS/m)

(Automatic temperature compensation at 25)

Cell constant: 0.1cm-2(10m-2)

Sensor input

1 piece

Water temperature

0~100

Temperature compensation

0~100

Accuracy

Repeatability±1.0% Full Scale

Analog output

4-20mA DC(resistance load less than 400Ω)

Conductivity and temperature outputs

(Ground is common connection)

Operating condition

Ambient temperature  0~40

Ambient humidity     5~80%

(There should be no dew condensation.)

Power supply

DC 24V±10% (exterior supply)

*External power supply and internal circuit are insulated electrically.

Power consumption

MAX 3W (Lower than DC24V/125mA)

Display

Not available

Dimensions

W45xH70xD109 mm (35mmDIN rail install type)

Usable sensor

R3/4 type

*7701-S300

*7701-S310

*7701-S600

R1/2 type

*7701-S400

*7701-S500

Others

It is possible to extend the cable up to a maximum 20 meters.

Use the cable specified by Techno Morioka.