Products

  • Home
  • Products
  • On-line Ions Analyzer
  • On-line Copper Ions Analyzer

ETi-9510

    Features
    ●Analysis Range: 0 ~ 6 ppm
    ●The patent design of both of quantitative sampling and dilution device require no maintenance and regular flow calibration.
    ●The use of large caliber sampling cup and sampling tube causes no pipe blockage.
    ●The sampling tube and drain tube with large aperture pneumatic valve, which prevents from the fouling or blockage and it may happened in traditional solenoid valve.
    ●While doing the measurement, first deduct the background absorbance to control the Solid Suspension influence.
    ●If the particulate matter concentration of water sample< 1% and particle size < 1mm, no need to install the filter.
    ●If the particulate matter concentration of water sample< 1% and particle size < 1mm, no need to install the filter.
    ●Auto regular calibration (the cycle time could be set by program).
    ●It can manually analyze the foreign samples for the person in laboratory to do the comparison conveniently.

内容说明

 

Model.

ETi-9510-Cu

Display Screen

7 inches of color touch screen

Analysis Range

0 ~ 6 ppm

pH Value of Water Sample

pH 4 ~ 12

Sample Temp.

5 ~ 45

Analysis Cycle Time

6 ~ 8 mins.

Accuracy

6 ppm ± 0.42 ppm.

Reproducibility

± 3 % full scale

Resolution

0.001 ppm

AutoCorrect Message and Calibration Record

V

Standby Time

0 ~ 7200 (sec)

Historical Data and Record

V

Alarm Message and Record

V

Liquid Level Detection

V

Auto Dosing

V