Products

  • Home
  • Products
  • PH Meter
  • ETi-6101/6121(Transmitter/Controller)

ETi-6101/6121(Transmitter / Controller)

    ■ Features
    ● Drinking water and groundwater monitor
    ● Industry wastewater monitor
    ● Chemical industry and water treatment
    ● Monitor the process with Fluoride and wastewater in water industry
    ● Pure water or high interference environment 6101D

内容说明

 Specification

Analysis Range

pH 0~14 or (pH -2~16 need to be scheduled in advance)

Accuracy

± 0.01 PH±0.2 

Value

0.01 PH

pH Electrode

A set of pH electrode is available

Temperature Compensation

The standard fitting is NTC10K temperature sensor, and Pt100 is selectable. Auto or manual temperature compensation is optional.

Electrode Response Time

< 60 sec

Calibration

Two-point manual calibration. Zero-point calibration with pH 7 or pH 6.86 calibration liquid, and it features auto buffer recognition.

Current Output

A set of current output 4 ~ 20 mA

Control Output

6101

None

6121

2 relay output with contact capacity (24VDC, 110/220 VAC) to startup alarm or dosing pump

Sample Temp.

0~100 °C

Applicable Environment

The case of wall-mounted ETi- 6101/6121-AL is made of aluminum alloy, which is waterproof completely and can be used outdoors.

Dimension

ETi- 6101/6121-AL 192 X 251 X 90

Power Supply

110/220 50/60Hz VAC